Baxter
CVE-2022-26390
MEDIUM
Severity by source
AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
Primary rating from NVD · only source for this CVE.
CVSS VectorNVD
Lifecycle Timeline
1DescriptionNVD
The Baxter Spectrum Wireless Battery Module (WBM) stores network credentials and PHI (only applicable to Spectrum IQ pumps using auto programming) in unencrypted form. An attacker with physical access to a device that hasn't had all data and settings erased may be able to extract sensitive information.
AnalysisAI
The Baxter Spectrum Wireless Battery Module (WBM) stores network credentials and PHI (only applicable to Spectrum IQ pumps using auto programming) in unencrypted form. Rated medium severity (CVSS 4.2), this vulnerability is no authentication required. No vendor patch available.
Technical ContextAI
This vulnerability is classified under CWE-311. The Baxter Spectrum Wireless Battery Module (WBM) stores network credentials and PHI (only applicable to Spectrum IQ pumps using auto programming) in unencrypted form. An attacker with physical access to a device that hasn't had all data and settings erased may be able to extract sensitive information. Affected products include: Baxter Spectrum Wireless Battery Module Firmware, Baxter Sigma Spectrum 35700Bax Firmware, Baxter Sigma Spectrum 35700Bax2 Firmware, Baxter Baxter Spectrum Iq 35700Bax3 Firmware.
RemediationAI
No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Apply vendor patches when available. Implement network segmentation and monitoring as interim mitigations.
In Connex health portal released before8/30/2024, SQL injection vulnerabilities were found that could have allowed an un
The impacted products, when configured to use SSO, are affected by an improper authentication vulnerability. Rated criti
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP servic
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BA
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensit
In Baxter Connex health portal released before 8/30/2024, an improper access control vulnerability has been found that c
Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensi
The Baxter Spectrum WBM is susceptible to format string attacks via application messaging. Rated high severity (CVSS 8.1
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit
Same weakness CWE-311 – Missing Encryption of Sensitive Data
View allSame technique Information Disclosure
View allShare
External POC / Exploit Code
Leaving vuln.today