Baxter
Monthly
In Baxter Connex health portal released before 8/30/2024, an improper access control vulnerability has been found that could allow an unauthenticated attacker to gain unauthorized access to Connex. Rated critical severity (CVSS 9.1), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
In Connex health portal released before8/30/2024, SQL injection vulnerabilities were found that could have allowed an unauthenticated attacker to gain unauthorized access to Connex portal's database. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
The Baxter Spectrum WBM does not perform mutual authentication with the gateway server host. Rated medium severity (CVSS 5.4), this vulnerability is no authentication required, low attack complexity. No vendor patch available.
The Baxter Spectrum WBM is susceptible to format string attacks via application messaging. Rated high severity (CVSS 8.1), this vulnerability is remotely exploitable, low attack complexity. No vendor patch available.
The Baxter Spectrum WBM (v16, v16D38) and Baxter Spectrum WBM (v17, v17D19, v20D29 to v20D32) when in superuser mode is susceptible to format string attacks via application messaging. Rated medium severity (CVSS 6.5), this vulnerability is remotely exploitable, low attack complexity. No vendor patch available.
The Baxter Spectrum Wireless Battery Module (WBM) stores network credentials and PHI (only applicable to Spectrum IQ pumps using auto programming) in unencrypted form. Rated medium severity (CVSS 4.2), this vulnerability is no authentication required. No vendor patch available.
The impacted products, when configured to use SSO, are affected by an improper authentication vulnerability. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on. Rated high severity (CVSS 7.5), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP service operating on the WBM remains operational until the WBM is rebooted. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to. Rated critical severity (CVSS 9.4), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered. Rated low severity (CVSS 2.4), this vulnerability is no authentication required, low attack complexity. No vendor patch available.
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS. Rated high severity (CVSS 7.5), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS. Rated high severity (CVSS 7.5), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings,. Rated medium severity (CVSS 4.9), this vulnerability is low attack complexity. No vendor patch available.
Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensitive information in an unencrypted database. Rated critical severity (CVSS 9.1), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical. Rated medium severity (CVSS 6.1), this vulnerability is no authentication required, low attack complexity. No vendor patch available.
Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and. Rated medium severity (CVSS 6.1), this vulnerability is low attack complexity. No vendor patch available.
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13,. Rated medium severity (CVSS 6.1), this vulnerability is no authentication required, low attack complexity. No vendor patch available.
Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems use cleartext messages to communicate order information with an order entry system. Rated high severity (CVSS 7.5), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
In Baxter Connex health portal released before 8/30/2024, an improper access control vulnerability has been found that could allow an unauthenticated attacker to gain unauthorized access to Connex. Rated critical severity (CVSS 9.1), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
In Connex health portal released before8/30/2024, SQL injection vulnerabilities were found that could have allowed an unauthenticated attacker to gain unauthorized access to Connex portal's database. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
The Baxter Spectrum WBM does not perform mutual authentication with the gateway server host. Rated medium severity (CVSS 5.4), this vulnerability is no authentication required, low attack complexity. No vendor patch available.
The Baxter Spectrum WBM is susceptible to format string attacks via application messaging. Rated high severity (CVSS 8.1), this vulnerability is remotely exploitable, low attack complexity. No vendor patch available.
The Baxter Spectrum WBM (v16, v16D38) and Baxter Spectrum WBM (v17, v17D19, v20D29 to v20D32) when in superuser mode is susceptible to format string attacks via application messaging. Rated medium severity (CVSS 6.5), this vulnerability is remotely exploitable, low attack complexity. No vendor patch available.
The Baxter Spectrum Wireless Battery Module (WBM) stores network credentials and PHI (only applicable to Spectrum IQ pumps using auto programming) in unencrypted form. Rated medium severity (CVSS 4.2), this vulnerability is no authentication required. No vendor patch available.
The impacted products, when configured to use SSO, are affected by an improper authentication vulnerability. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on. Rated high severity (CVSS 7.5), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP service operating on the WBM remains operational until the WBM is rebooted. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to. Rated critical severity (CVSS 9.4), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered. Rated low severity (CVSS 2.4), this vulnerability is no authentication required, low attack complexity. No vendor patch available.
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS. Rated high severity (CVSS 7.5), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS. Rated high severity (CVSS 7.5), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings,. Rated medium severity (CVSS 4.9), this vulnerability is low attack complexity. No vendor patch available.
Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensitive information in an unencrypted database. Rated critical severity (CVSS 9.1), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical. Rated medium severity (CVSS 6.1), this vulnerability is no authentication required, low attack complexity. No vendor patch available.
Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and. Rated medium severity (CVSS 6.1), this vulnerability is low attack complexity. No vendor patch available.
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13,. Rated medium severity (CVSS 6.1), this vulnerability is no authentication required, low attack complexity. No vendor patch available.
Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems use cleartext messages to communicate order information with an order entry system. Rated high severity (CVSS 7.5), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.