Skip to main content

Atlan A350 EUVDEUVD-2021-34845

| CVE-2021-4479 MEDIUM
Improper Validation of Syntactic Correctness of Input (CWE-1286)
2026-06-02 VulnCheck GHSA-744p-v485-p6h6
6.3
CVSS 4.0 · NVD
Share

Severity by source

NVD PRIMARY
6.3 MEDIUM
CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

Primary rating from NVD · only source for this CVE.

CVSS VectorNVD

CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Attack Vector
Network
Attack Complexity
High
Privileges Required
None
User Interaction
None
Scope
X

Lifecycle Timeline

3
Patch available
Jun 02, 2026 - 21:02 EUVD
CVSS changed
Jun 02, 2026 - 20:22 NVD
4.0 (MEDIUM) 6.3 (MEDIUM)
CVE Published
Jun 02, 2026 - 19:27 cve.org
MEDIUM 6.3

DescriptionCVE.org

Dräger Atlan A350 software versions 1.00 through 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.

AnalysisAI

Dräger Atlan A350 software versions 1.00 through 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted. Rated medium severity (CVSS 6.3), this vulnerability is remotely exploitable, no authentication required.

Technical ContextAI

This vulnerability is classified under CWE-1286. Dräger Atlan A350 software versions 1.00 through 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves. Affected products include: Dräger Atlan A350. Version information: through 1.01.

RemediationAI

A vendor patch is available. Apply the latest security update as soon as possible. Apply vendor patches when available. Implement network segmentation and monitoring as interim mitigations.