Skip to main content

Perseus A500 EUVDEUVD-2019-20159

| CVE-2019-25723 MEDIUM
Improper Validation of Syntactic Correctness of Input (CWE-1286)
2026-06-02 VulnCheck GHSA-9p7r-jmp9-7hw2
6.3
CVSS 4.0 · NVD
Share

Severity by source

NVD PRIMARY
6.3 MEDIUM
CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

Primary rating from NVD · only source for this CVE.

CVSS VectorNVD

CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Attack Vector
Network
Attack Complexity
High
Privileges Required
None
User Interaction
None
Scope
X

Lifecycle Timeline

2
CVSS changed
Jun 02, 2026 - 20:22 NVD
4.0 (MEDIUM) 6.3 (MEDIUM)
CVE Published
Jun 02, 2026 - 19:00 cve.org
MEDIUM 6.3

DescriptionCVE.org

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

AnalysisAI

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted. Rated medium severity (CVSS 6.3), this vulnerability is remotely exploitable, no authentication required. No vendor patch available.

Technical ContextAI

This vulnerability is classified under CWE-1286. Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes. Affected products include: Dräger Perseus A500. Version information: through 2.02.

RemediationAI

No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Apply vendor patches when available. Implement network segmentation and monitoring as interim mitigations.

Share

EUVD-2019-20159 vulnerability details – vuln.today

This site uses cookies essential for authentication and security. No tracking or analytics cookies are used. Privacy Policy