Pyxis Medstation Es Firmware
CVE-2020-10598
MEDIUM
Severity by source
AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
Primary rating from NVD · only source for this CVE.
CVSS VectorNVD
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
Lifecycle Timeline
1DescriptionNVD
In BD Pyxis MedStation ES System v1.6.1 and Pyxis Anesthesia (PAS) ES System v1.6.1, a restricted desktop environment escape vulnerability exists in the kiosk mode functionality of affected devices. Specially crafted inputs could allow the user to escape the restricted environment, resulting in access to sensitive data.
AnalysisAI
In BD Pyxis MedStation ES System v1.6.1 and Pyxis Anesthesia (PAS) ES System v1.6.1, a restricted desktop environment escape vulnerability exists in the kiosk mode functionality of affected devices. Rated medium severity (CVSS 6.1), this vulnerability is no authentication required, low attack complexity. No vendor patch available.
Technical ContextAI
This vulnerability is classified under CWE-693. In BD Pyxis MedStation ES System v1.6.1 and Pyxis Anesthesia (PAS) ES System v1.6.1, a restricted desktop environment escape vulnerability exists in the kiosk mode functionality of affected devices. Specially crafted inputs could allow the user to escape the restricted environment, resulting in access to sensitive data. Affected products include: Bd Pyxis Medstation Es Firmware, Bd Pyxis Anesthesia Station Es Firmware.
RemediationAI
No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Apply vendor patches when available. Implement network segmentation and monitoring as interim mitigations.
More in Pyxis Medstation Es Firmware
View allSame weakness CWE-693 – Protection Mechanism Failure
View allSame technique Information Disclosure
View allShare