Scroll Medical Air Systems Firmware
CVE-2018-7518
CRITICAL
Severity by source
AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Primary rating from NVD · only source for this CVE.
CVSS VectorNVD
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Lifecycle Timeline
1DescriptionNVD
In TotalAlert Web Application in BeaconMedaes Scroll Medical Air Systems prior to v4107600010.23, an attacker with network access to the integrated web server could retrieve default or user defined credentials stored and transmitted in an insecure manner.
AnalysisAI
In TotalAlert Web Application in BeaconMedaes Scroll Medical Air Systems prior to v4107600010.23, an attacker with network access to the integrated web server could retrieve default or user defined. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
Technical ContextAI
This vulnerability is classified as Insufficiently Protected Credentials (CWE-522), which allows attackers to obtain user credentials due to weak protection mechanisms. In TotalAlert Web Application in BeaconMedaes Scroll Medical Air Systems prior to v4107600010.23, an attacker with network access to the integrated web server could retrieve default or user defined credentials stored and transmitted in an insecure manner. Affected products include: Beaconmedaes Scroll Medical Air Systems Firmware.
RemediationAI
No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Hash passwords with strong algorithms (bcrypt, argon2), encrypt credentials in transit and at rest, never log credentials.
In the web application in BeaconMedaes TotalAlert Scroll Medical Air Systems running software versions prior to 41076000
In TotalAlert Web Application in BeaconMedaes Scroll Medical Air Systems prior to v4107600010.23, by accessing a specifi
Same weakness CWE-522 – Insufficiently Protected Credentials
View allSame technique Information Disclosure