2090 Carelink Programmer Firmware
CVE-2018-10596
HIGH
Vulnerability Assessment
| Weakness |
CWE
ASVS 5.0
V12 - Secure Communication Improper restriction of communication channel
|
| Impact | CVSS score of 7.1 (HIGH). Attack characteristics: network attack vector, high complexity, low privileges required, user interaction required. |
| Exploit ScenarioDerived | This vulnerability can be exploited remotely over the network. Successful exploitation could compromise the high-severity affected system. |
Get an e-mail when this CVE is exploited, patched or re-scored. Free account.
Alert meAnalysisAI
Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. Rated high severity (CVSS 7.1), this vulnerability is remotely exploitable. No vendor patch available.
RemediationAI
No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Apply vendor patches when available. Implement network segmentation and monitoring as interim mitigations.
Affects your stack
Find out whether the stacks and servers you run are on an affected version.
Checking this CVE against your own stacks and servers is part of the Basic plan and up.
Technical ContextAI · show details
This vulnerability is classified under CWE-923. Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications. Affected products include: Medtronic 2090 Carelink Programmer Firmware.
Know when 2090 Carelink Programmer Firmware gets its next vulnerability
- Watch products and get an email when something new is published
- Check this CVE against your own stack and servers
- Exploit and PoC references, threat intelligence from 22 sources
No credit card · 7-day full trial
Severity by source
AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H
Primary rating from Vendor (hq) · only source for this CVE.
CVSS VectorVendor: hq
DescriptionCVE.org
Medtronic 2090 CareLink Programmer
uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.
Threat Intelligence Overview
How this decision was reached
The engine reads each signal below and applies a fixed list of rules from top to bottom. The first rule that matches sets the recommended action – like a statement, every line is shown.
What every action level means
SSVC AssessmentENISA
Priority Score
Attack Technique
Information disclosure occurs when an application unintentionally exposes sensitive data that aids attackers in reconnaissance or directly compromises security.
Affected Products (CPE)CVE.org
Medtronic 2090 CareLink Programmer uses a per-product username and password that is stored in a recoverable format. Rate
Medtronic 2090 CareLink Programmer’s software deployment network contains a directory traversal vulnerability that could
NVD
CVE.org
EPSS
GitHub PoC
Exploit-DB