Skip to main content

Cms8000 Firmware CVE-2022-38453

MEDIUM
Active Debug Code (CWE-489)
2022-09-13 ics-cert@hq.dhs.gov
4.4
CVSS 3.1 · NVD
Share

Severity by source

NVD PRIMARY
4.4 MEDIUM
AV:L/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:N

Primary rating from NVD · only source for this CVE.

CVSS VectorNVD

CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:L/I:L/A:N
Attack Vector
Local
Attack Complexity
Low
Privileges Required
Low
User Interaction
None
Scope
Unchanged
Confidentiality
Low
Integrity
Low
Availability
None

Lifecycle Timeline

1
CVE Published
Sep 13, 2022 - 15:15 nvd
MEDIUM 4.4

DescriptionNVD

Multiple binary application files on the CMS8000 device are compiled with 'not stripped' and 'debug_info' compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities.

AnalysisAI

Multiple binary application files on the CMS8000 device are compiled with 'not stripped' and 'debug_info' compilation settings. Rated medium severity (CVSS 4.4), this vulnerability is low attack complexity. No vendor patch available.

Technical ContextAI

This vulnerability is classified under CWE-489. Multiple binary application files on the CMS8000 device are compiled with 'not stripped' and 'debug_info' compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities. Affected products include: Contechealth Cms8000 Firmware.

RemediationAI

No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Apply vendor patches when available. Implement network segmentation and monitoring as interim mitigations.

Share

CVE-2022-38453 vulnerability details – vuln.today

This site uses cookies essential for authentication and security. No tracking or analytics cookies are used. Privacy Policy