Mycarelink Smart Model 25000 Firmware
CVE-2020-25183
HIGH
Severity by source
AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Primary rating from NVD · only source for this CVE.
CVSS VectorNVD
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Lifecycle Timeline
1DescriptionNVD
Medtronic MyCareLink Smart 25000 contains
an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.
AnalysisAI
Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile. Rated high severity (CVSS 8.8), this vulnerability is no authentication required, low attack complexity. No vendor patch available.
Technical ContextAI
This vulnerability is classified as Improper Authentication (CWE-287), which allows attackers to bypass authentication mechanisms to gain unauthorized access. Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication. Affected products include: Medtronic Mycarelink Smart Model 25000 Firmware.
RemediationAI
No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Implement multi-factor authentication, enforce strong password policies, use proven authentication frameworks.
Medtronic MyCareLink Smart 25000 is vulnerable when an authenticated attacker runs a debug command, which can be sent to
Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update syste
Same weakness CWE-287 – Improper Authentication
View allSame technique Authentication Bypass