Skip to main content

Mycarelink Smart Model 25000 Firmware CVE-2020-25183

HIGH
Improper Authentication (CWE-287)
2020-12-14 ics-cert@hq.dhs.gov
8.8
CVSS 3.1 · NVD
Share

Severity by source

NVD PRIMARY
8.8 HIGH
AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Primary rating from NVD · only source for this CVE.

CVSS VectorNVD

CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Attack Vector
Adjacent
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
High
Availability
High

Lifecycle Timeline

1
CVE Published
Dec 14, 2020 - 20:15 nvd
HIGH 8.8

DescriptionNVD

Medtronic MyCareLink Smart 25000 contains

an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.

AnalysisAI

Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile. Rated high severity (CVSS 8.8), this vulnerability is no authentication required, low attack complexity. No vendor patch available.

Technical ContextAI

This vulnerability is classified as Improper Authentication (CWE-287), which allows attackers to bypass authentication mechanisms to gain unauthorized access. Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication. Affected products include: Medtronic Mycarelink Smart Model 25000 Firmware.

RemediationAI

No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Implement multi-factor authentication, enforce strong password policies, use proven authentication frameworks.

Share

CVE-2020-25183 vulnerability details – vuln.today

This site uses cookies essential for authentication and security. No tracking or analytics cookies are used. Privacy Policy