Vulnerability Assessment
| Weakness |
CWE
OWASP Top 10:2025
A06:2025 - Insecure Design
ASVS 5.0
V2 - Validation and Business Logic Resource consumption limits
MITRE ATT&CK
1 technique across 1 tactic: Impact The technique list with ATT&CK links and the kill-chain view is part of the Basic plan and up. |
| Impact | CVSS score of 7.1 (HIGH). Attack characteristics: adjacent network attack vector, low complexity, no privileges required, no user interaction not required. EPSS exploitation probability: 0.02%. |
| Exploit ScenarioDerived | An attacker could exploit this vulnerability to cause denial of service by exhausting system resources. No authentication is required to exploit this vulnerability. |
Get an e-mail when this CVE is exploited, patched or re-scored. Free account.
Alert meAnalysisAI
Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or. Rated high severity (CVSS 7.1), this vulnerability is no authentication required, low attack complexity. No vendor patch available.
RemediationAI
No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Implement rate limiting, set resource quotas, validate input sizes, use timeouts.
Affected VersionsENISA
Affects your stack
Find out whether the stacks and servers you run are on an affected version.
Checking this CVE against your own stacks and servers is part of the Basic plan and up.
Technical ContextAI · show details
This vulnerability is classified as Uncontrolled Resource Consumption (CWE-400), which allows attackers to cause denial of service by exhausting system resources. Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered. Affected products include: Dräger Infinity M300.
Know when Infinity M300 gets its next vulnerability
- Watch products and get an email when something new is published
- Check this CVE against your own stack and servers
- Exploit and PoC references, threat intelligence from 22 sources
No credit card · 7-day full trial
Severity by source
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
Primary rating from Vendor (VulnCheck) · only source for this CVE.
CVSS VectorVendor: VulnCheck
Lifecycle Timeline
3DescriptionCVE.org
Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.
Threat Intelligence Overview
How this decision was reached
The engine reads each signal below and applies a fixed list of rules from top to bottom. The first rule that matches sets the recommended action – like a statement, every line is shown.
What every action level means
SSVC AssessmentENISA
Priority Score
Attack Technique
Denial of Service attacks render applications or systems unavailable by overwhelming resources or triggering failure conditions.
EUVD-2019-20160
GHSA-9492-c6v8-vj2h
NVD
CVE.org
EPSS
Vulnrichment
Exploit-DB