24950 Mycarelink Monitor Firmware
CVE-2018-8870
MEDIUM
Vulnerability Assessment
| Weakness |
CWE
OWASP Top 10:2025
A07:2025 - Authentication Failures
ASVS 5.0
V6 - Authentication No hard-coded passwords
MITRE ATT&CK
1 technique across 1 tactic: Persistence The technique list with ATT&CK links and the kill-chain view is part of the Basic plan and up. |
| Impact | CVSS score of 6.4 (MEDIUM). Attack characteristics: physical attack vector, high complexity, no privileges required, no user interaction not required. |
| Exploit ScenarioDerived | Successful exploitation could compromise the medium-severity affected system. |
Get an e-mail when this CVE is exploited, patched or re-scored. Free account.
Alert meAnalysisAI
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. Rated medium severity (CVSS 6.4), this vulnerability is no authentication required. No vendor patch available.
RemediationAI
No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Apply vendor patches when available. Implement network segmentation and monitoring as interim mitigations.
Affects your stack
Find out whether the stacks and servers you run are on an affected version.
Checking this CVE against your own stacks and servers is part of the Basic plan and up.
Technical ContextAI · show details
This vulnerability is classified under CWE-259. Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system. Affected products include: Medtronic 24950 Mycarelink Monitor Firmware, Medtronic 24952 Mycarelink Monitor Firmware.
Know when 24950 Mycarelink Monitor Firmware gets its next vulnerability
- Watch products and get an email when something new is published
- Check this CVE against your own stack and servers
- Exploit and PoC references, threat intelligence from 22 sources
No credit card · 7-day full trial
Severity by source
AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
Primary rating from Vendor (hq) · only source for this CVE.
CVSS VectorVendor: hq
DescriptionCVE.org
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.
Threat Intelligence Overview
How this decision was reached
The engine reads each signal below and applies a fixed list of rules from top to bottom. The first rule that matches sets the recommended action – like a statement, every line is shown.
What every action level means
SSVC AssessmentENISA
Priority Score
Attack Technique
Authentication bypass attacks exploit flaws in the verification mechanisms that control access to systems and applications.
NVD
CVE.org
EPSS
GitHub PoC
Exploit-DB