CVE-2020-16220 MEDIUM This Month

In Patient Information Center iX (PICiX) Versions C.02, C.03, PerformanceBridge Focal Point Version A.01, the product receives input that is expected to be well-formed (i.e., to comply with a certain. Rated medium severity (CVSS 4.3), this vulnerability is no authentication required, low attack complexity. No vendor patch available.

Denial Of Service Patient Information Center Ix Performancebridge Focal Point
NVD
CVSS 3.1
4.3
EPSS
0.3%
CVE-2020-16216 MEDIUM This Month

In IntelliVue patient monitors MX100, MX400-550, MX600, MX700, MX750, MX800, MX850, MP2-MP90, and IntelliVue X2 and X3 Versions N and prior, the product receives input or data but does not validate. Rated medium severity (CVSS 6.5), this vulnerability is no authentication required, low attack complexity. No vendor patch available.

Information Disclosure Patient Information Center Ix Performancebridge Focal Point Intellivue Mp2 Mp90 Firmware Intellivue Mx100 Firmware +9
NVD
CVSS 3.1
6.5
EPSS
0.7%
CVE-2020-16228 MEDIUM This Month

In Patient Information Center iX (PICiX) Versions C.02 and C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N. Rated medium severity (CVSS 6.4), this vulnerability is low attack complexity. No vendor patch available.

Information Disclosure Patient Information Center Ix Performancebridge Focal Point Intellivue Mp2 Mp90 Firmware Intellivue Mx100 Firmware +9
NVD
CVSS 3.1
6.4
EPSS
0.4%
CVE-2020-16222 HIGH This Week

In Patient Information Center iX (PICiX) Version B.02, C.02, C.03, and PerformanceBridge Focal Point Version A.01, when an actor claims to have a given identity, the software does not prove or. Rated high severity (CVSS 8.8), this vulnerability is no authentication required, low attack complexity. No vendor patch available.

Authentication Bypass Patient Information Center Ix Performancebridge Focal Point
NVD
CVSS 3.1
8.8
EPSS
0.6%
EPSS 0% CVSS 4.3
MEDIUM This Month

In Patient Information Center iX (PICiX) Versions C.02, C.03, PerformanceBridge Focal Point Version A.01, the product receives input that is expected to be well-formed (i.e., to comply with a certain. Rated medium severity (CVSS 4.3), this vulnerability is no authentication required, low attack complexity. No vendor patch available.

Denial Of Service Patient Information Center Ix Performancebridge Focal Point
NVD
EPSS 1% CVSS 6.5
MEDIUM This Month

In IntelliVue patient monitors MX100, MX400-550, MX600, MX700, MX750, MX800, MX850, MP2-MP90, and IntelliVue X2 and X3 Versions N and prior, the product receives input or data but does not validate. Rated medium severity (CVSS 6.5), this vulnerability is no authentication required, low attack complexity. No vendor patch available.

Information Disclosure Patient Information Center Ix Performancebridge Focal Point +11
NVD
EPSS 0% CVSS 6.4
MEDIUM This Month

In Patient Information Center iX (PICiX) Versions C.02 and C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N. Rated medium severity (CVSS 6.4), this vulnerability is low attack complexity. No vendor patch available.

Information Disclosure Patient Information Center Ix Performancebridge Focal Point +11
NVD
EPSS 1% CVSS 8.8
HIGH This Week

In Patient Information Center iX (PICiX) Version B.02, C.02, C.03, and PerformanceBridge Focal Point Version A.01, when an actor claims to have a given identity, the software does not prove or. Rated high severity (CVSS 8.8), this vulnerability is no authentication required, low attack complexity. No vendor patch available.

Authentication Bypass Patient Information Center Ix Performancebridge Focal Point
NVD