Skip to main content

CGM CLININET EUVDEUVD-2025-211177

| CVE-2025-30046 CRITICAL
Execution with Unnecessary Privileges (CWE-250)
2026-10-23 CERT-PL GHSA-q872-9p8m-4vww
9.0
CVSS 4.0 · Vendor: CERT-PL
View

Attack ChainAIDerived

Hypothetical attack flow derived from CVE metadata – not a confirmed exploit chain

  1. Access
    Gain adjacent-network foothold
  2. Delivery
    Send crafted HL7 data to listener
  3. Exploit
    Trigger flaw in root HL7 process
  4. Execution
    Execute arbitrary code as root
  5. Impact
    Full compromise of application server

Vulnerability Assessment

Weakness
CWE
ASVS 5.0
V8 - Authorization Unnecessary privilege execution
Exploitation Conditions

Exploitation requires network reachability to the CGM CLININET HL7 communication processes from an ADJACENT network segment (AV:A) - these processes run with root privileges and parse untrusted data received from external/integrated systems, and no authentication is required (PR:N). The concrete prerequisite is therefore that the attacker can deliver data to an exposed HL7 interface/listener; the limiting factors are that the attacker must already be on an adjacent (not arbitrary remote) network, and that CVSS encodes an additional attack requirement (AT:P) whose specific trigger condition is not described in the available sources. Affected versions are stated up to 2026.MS3, with later versions possibly also affected per the vendor/reporter.

Impact

This is a genuine high-priority issue rather than an inflated-CVSS artifact. The CVSS 4.0 base score of 9.0 is driven by unauthenticated access (PR:N), low complexity (AC:L), no user interaction (UI:N), and total compromise of both the vulnerable and subsequent systems (VC/VI/VA:H and SC/SI/SA:H), reflecting that the affected HL7 processes run as root (CWE-250, execution with unnecessary privilege). The vuln.today decision label is act_now and the finding was reported by CERT-PL, a national CERT, which raises confidence that this is a real, prioritised defect. The single meaningful mitigating signal in the vector is the attack vector: AV:A (adjacent network) means the attacker must already have a foothold on a network segment that can reach the HL7 communication processes, not arbitrary Internet reachability. AT:P indicates the CVSS author judged that some additional attack requirement exists, but the available sources do not describe what that requirement or the trigger mechanism actually is, so it should not be read as a strong barrier. Key risk signals are MISSING and should be stated explicitly: no EPSS exploitation-probability score is provided, the CVE is not listed in CISA KEV, and no proof-of-concept or exploit code is referenced. Absence of EPSS/KEV/POC lowers the evidence for active or imminent mass exploitation but does not reduce the technical severity; given act_now and a national-CERT report, treat this as a patch-urgently item for any CGM CLININET deployment.

Exploit ScenarioDerived

An attacker who has gained a position on a network segment adjacent to the CGM CLININET application server (for example a compromised device on the hospital/medical-device VLAN used for HL7 integration) sends crafted data to one of the root-privileged HL7 communication processes, which processes it without authentication and is driven to execute attacker-controlled code as root. No public exploit code is identified at time of analysis, so this scenario is inferred from the description and CVSS vector rather than a published PoC.

Act Now Prioritize within 24-48h. Why: CRITICAL severity.
No patch listed yet 0.3% exploitation probability (EPSS)

Get an e-mail when this CVE is exploited, patched or re-scored. Free account.

Alert me

AnalysisAIv1 – Oct 23, 2026

Remote code execution with root privileges affects CGM CLININET application servers through version 2026.MS3, where HL7 communication processes run as root and parse untrusted data from external systems. An unauthenticated attacker on an adjacent network segment who can deliver data to an exposed HL7 listener can achieve code execution with root privileges; the additional attack requirement encoded as AT:P is not described in the available sources. No public exploit was identified at time of analysis, but the issue is a genuine high-priority item prioritized act_now because of the root-level impact and unauthenticated adjacent-network path.

RemediationAI

No exact fixed version is listed in the provided data; the reporter states impact up to 2026.MS3 with later versions possibly affected, so contact CGM support and monitor the CERT.PL advisory at https://cert.pl/en/posts/2026/10/CVE-2025-30043/ and the CGM product page at https://www.cgm.com/pol_pl/products/szpital/cgm-clininet.html for a patched release. Until a fixed build is confirmed, restrict network reachability to the CGM CLININET HL7 communication processes and listeners so that only trusted external systems on tightly controlled adjacent segments can deliver data; this directly addresses the adjacent-network delivery prerequisite but may disrupt integrations and does not remediate the root-privilege parsing flaw. Avoid expanding HL7 interface exposure while awaiting the vendor fix.

Affected VersionsENISA

CGM CLININET 0 ≤2026.MS3

Affected Packages

euvd cgm_clininet no fix

Source: OSV (Open Source Vulnerabilities)

Affects your stack

Affects 1 package in euvd

Checking this CVE against your own stacks and servers is part of the Basic plan and up.

Technical ContextAI · show details

CGM CLININET is a healthcare application server, and the affected component is its HL7 communication layer, which exchanges messages with external and integrated systems. Multiple processes handling HL7 traffic execute with root privileges while processing untrusted data received from those external systems, matching CWE-250 (Execution with Unnecessary Privileges). The CVSS vector confirms adjacent-network access (AV:A), no privileges required (PR:N), no user interaction (UI:N), high confidentiality, integrity, and availability impact, and a scope change (S:C); it also includes an additional attack requirement (AT:P) whose specific trigger condition is not described in the available evidence. The CPE cpe:2.3:a:cgm:cgm_clininet:*:*:*:*:*:*:*:* identifies CGM CLININET as the affected product, with reported impact through version 2026.MS3 and possible impact on later versions.

Know when Cgm Clininet gets its next vulnerability

  • Watch products and get an email when something new is published
  • Check this CVE against your own stack and servers
  • Exploit and PoC references, threat intelligence from 22 sources

No credit card · 7-day full trial

Severity by source

Vendor (CERT-PL) PRIMARY 9.0 CRITICAL
CVSS:4.0/AV:A/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:H/SI:H/SA:H
vuln.today AI 9.6 CRITICAL

Unauthenticated adjacent attacker sends data to a root-level HL7 process (AV:A/PR:N), and code execution as root beyond the vulnerable component yields scope change and full C/I/A impact.

3.1 AV:A/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
4.0 AV:A/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:H/SI:H/SA:H

Primary rating from Vendor (CERT-PL).

CVSS VectorVendor: CERT-PL

Attack Vector
Adjacent
Attack Complexity
Low
Attack Requirements
Present
Privileges Required
None
User Interaction
None
Vulnerable system: Confidentiality
High
Vulnerable system: Integrity
High
Vulnerable system: Availability
High
Subsequent system: Confidentiality
High
Subsequent system: Integrity
High
Subsequent system: Availability
High

DescriptionCVE.org

In the CGM CLININET application server, multiple processes responsible for HL7 communication with external systems run with root privileges while processing untrusted data received from those external systems. An attacker in adjacent network able to send data to these HL7 communication processes can achieve remote code execution with root privileges.

The vulnerability was confirmed to affect versions up to 2026.MS3 but later versions may also be vulnerable.

Executive ViewAI

Why It Matters

CVE-2025-30046 is a critical remote code execution vulnerability in CGM CLININET application servers (through 2026.MS3) that allows an unauthenticated attacker on an adjacent network segment to gain root-level control by sending malicious data to exposed HL7 listeners. This puts patient data confidentiality, integrity, and availability at severe risk, potentially disrupting clinical operations and violating HIPAA and GDPR obligations, which demands immediate leadership attention.

Business Impact: Severe
Compensating Controls

No vendor-released patch identified at time of analysis. Interim mitigations include: isolating affected servers from untrusted networks, restricting HL7 listener access to trusted IP addresses via firewall rules, and monitoring for anomalous HL7 traffic.

Threat Intelligence Overview

EPSS Score
0.3%
Top 83% exploitable
Prevalence
Low
industry-specific hospital information system, regional healthcare footprint
Social Media
2
Sightings
Data Quality
5 sources
Confidence: Medium

How this decision was reached

The engine reads each signal below and applies a fixed list of rules from top to bottom. The first rule that matches sets the recommended action – like a statement, every line is shown.

SignalValue
Severity CRITICAL
CISA KEV No
Public POC No
EPSS 0.3%
Patch available No
Active threat No
Deciding rule
CRITICAL severity
What every action level means
Emergency Immediate action required. Criteria: CRITICAL + in CISA KEV (actively exploited) without patch, OR CRITICAL + EPSS > 30% without patch, OR CRITICAL + active threat signals without patch.
Act Now Prioritize within 24-48h. Criteria: in KEV with patch available, OR CRITICAL + public POC, OR HIGH + EPSS > 10%, OR CRITICAL + active threat, OR any CRITICAL severity.
This Week Schedule remediation this week. Criteria: HIGH severity (CVSS 7.0-8.9). Patch within normal change window.
This Month Plan within 30 days. Criteria: MEDIUM severity (CVSS 4.0-6.9), lower exploitation probability. Include in regular patch cycle.
Monitor Track and reassess. Criteria: LOW severity (CVSS 0.1-3.9), minimal exploitation risk. No immediate action needed.

SSVC AssessmentENISA

SSVC Decision· derived
Scheduled
Exploitation· derived
None
System Exposure· derived
Controlled
Automatable· derived
No
Human Impact
Very High

Priority Score

45
Low Medium High Critical
KEV: 0
EPSS: 0
CVSS: +45
PoC: 0
Weaponized: 0
Threat intel: 0
Gov advisory: 0

Attack Technique

Privilege escalation occurs when an attacker leverages flaws in access control mechanisms to gain permissions beyond what they were originally granted.

Remote Code Execution represents the critical moment when an attacker successfully runs arbitrary code on a target system without physical access.

CVE-2025-30047 CRITICAL
9.4 Oct 23

Remote code execution in CGM CLININET lets unauthenticated attackers on the same local network execute arbitrary code as

CVE-2025-30050 CRITICAL
9.4 Oct 23

Code injection in the RunCGI.pm component of CGM CLININET allows an attacker positioned on an adjacent network segment t

CVE-2025-30054 CRITICAL
9.3 Oct 23

Local privilege escalation in CGM CLININET allows anyone who already holds the low-privileged uhcapache account to rewri

CVE-2025-30052 HIGH
8.8 Oct 23

Authentication bypass in the CGM CLININET rsyncd service allows an attacker who can spoof an allow-listed source IP addr

CVE-2025-30045 HIGH
8.6 Oct 23

Code injection in the ConvExpr function of CGM CLININET allows a high-privileged, authenticated user to execute arbitrar

CVE-2025-30049 HIGH
8.2 Oct 23

CGM CLININET stores user passwords as unsalted MD5 hashes, so once a credential store is obtained the passwords can be r

CVE-2025-30043 HIGH
7.1 Oct 23

CGM CLININET exposes its /DocManagment endpoint without an authentication check, allowing unauthenticated attackers on a

CVE-2025-30053 MEDIUM
6.9 Oct 23

The cgm-updater process shipped with CGM CLININET exposes a network listener that performs no authentication, so anythin

CVE-2025-30051 MEDIUM
6.8 Oct 23

CGM CLININET before 2026.MS3 exposes production database and RabbitMQ broker credentials to any local account because th

CVE-2025-58403 MEDIUM
5.1 Oct 23

Reflected cross-site scripting in CGM CLININET (confirmed through version 2026.MS3, with later releases possibly affecte

This site uses cookies essential for authentication and security. No tracking or analytics cookies are used. Privacy Policy