There CVE-2024-9832
CRITICALSeverity by source
AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Primary rating from Vendor (baxter) · only source for this CVE.
CVSS VectorVendor: baxter
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Lifecycle Timeline
1DescriptionCVE.org
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
AnalysisAI
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. Rated critical severity (CVSS 9.3), this vulnerability is no authentication required, low attack complexity. No vendor patch available.
Technical ContextAI
This vulnerability is classified under CWE-307. There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Affected ProductsAI
See vendor advisory for affected versions.
RemediationAI
No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Apply vendor patches when available. Implement network segmentation and monitoring as interim mitigations.