Skip to main content

Interstim X Clinician CVE-2023-25931

MEDIUM
Unverified Password Change (CWE-620)
2023-03-01 security@medtronic.com
6.8
CVSS 3.1 · NVD
Share

Severity by source

NVD PRIMARY
6.8 MEDIUM
AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Primary rating from NVD · only source for this CVE.

CVSS VectorNVD

CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Attack Vector
Physical
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
High
Availability
High

Lifecycle Timeline

1
CVE Published
Mar 01, 2023 - 19:15 nvd
MEDIUM 6.8

DescriptionNVD

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

AnalysisAI

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Rated medium severity (CVSS 6.8), this vulnerability is no authentication required, low attack complexity. No vendor patch available.

Technical ContextAI

This vulnerability is classified under CWE-620. Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer. Affected products include: Medtronic Interstim X Clinician, Medtronic Micro Clinician.

RemediationAI

No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Apply vendor patches when available. Implement network segmentation and monitoring as interim mitigations.