Hmi 3 Control Panel Firmware
CVE-2021-37167
CRITICAL
Vulnerability Assessment
| Weakness |
CWE
OWASP Top 10:2025
A06:2025 - Insecure Design
ASVS 5.0
V8 - Authorization Least privilege enforcement
MITRE ATT&CK
1 technique across 1 tactic: Privilege Escalation The technique list with ATT&CK links and the kill-chain view is part of the Basic plan and up. |
| Impact | CVSS score of 9.8 (CRITICAL). Attack characteristics: network attack vector, low complexity, no privileges required, no user interaction not required. |
| Exploit ScenarioDerived | A remote attacker could escalate privileges to gain unauthorized elevated access by sending crafted requests to the affected application. No authentication is required to exploit this vulnerability. |
Get an e-mail when this CVE is exploited, patched or re-scored. Free account.
Alert meAnalysisAI
An insecure permissions issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.
RemediationAI
No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Apply principle of least privilege, validate privilege transitions, implement proper role separation.
Affects your stack
Find out whether the stacks and servers you run are on an affected version.
Checking this CVE against your own stacks and servers is part of the Basic plan and up.
Technical ContextAI · show details
This vulnerability is classified as Improper Privilege Management (CWE-269), which allows attackers to escalate privileges to gain unauthorized elevated access. An insecure permissions issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. A user logged in using the default credentials can gain root access to the device, which provides permissions for all of the functionality of the device. Affected products include: Swisslog-Healthcare Hmi-3 Control Panel Firmware.
Know when Hmi 3 Control Panel Firmware gets its next vulnerability
- Watch products and get an email when something new is published
- Check this CVE against your own stack and servers
- Exploit and PoC references, threat intelligence from 22 sources
No credit card · 7-day full trial
Severity by source
AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Primary rating from NVD · only source for this CVE.
CVSS VectorNVD
DescriptionNVD
An insecure permissions issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. A user logged in using the default credentials can gain root access to the device, which provides permissions for all of the functionality of the device.
Threat Intelligence Overview
How this decision was reached
The engine reads each signal below and applies a fixed list of rules from top to bottom. The first rule that matches sets the recommended action – like a statement, every line is shown.
What every action level means
SSVC AssessmentENISA
Priority Score
Attack Technique
Privilege escalation occurs when an attacker leverages flaws in access control mechanisms to gain permissions beyond what they were originally granted.
Affected Products (CPE)NVD
More in Hmi 3 Control Panel Firmware
View allA buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released ver
An insecure permissions issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus operated by released ver
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released ver
A buffer overflow issue was discovered in the HMI3 Control Panel contained within the Swisslog Healthcare Nexus Panel, o
A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released ver
A buffer overflow issue leading to denial of service was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus P