Skip to main content

Severity by source

NVD PRIMARY
6.5 MEDIUM
AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N

Primary rating from NVD · only source for this CVE.

CVSS VectorNVD

CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
Attack Vector
Adjacent
Attack Complexity
Low
Privileges Required
None
User Interaction
None
Scope
Unchanged
Confidentiality
None
Integrity
High
Availability
None

Lifecycle Timeline

1
CVE Published
Mar 25, 2019 - 22:29 nvd
MEDIUM 6.5

DescriptionNVD

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

AnalysisAI

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia. Rated medium severity (CVSS 6.5), this vulnerability is no authentication required, low attack complexity. No vendor patch available.

Technical ContextAI

This vulnerability is classified under CWE-284. The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device. Affected products include: Medtronic Mycarelink Monitor Firmware, Medtronic Carelink Monitor Firmware, Medtronic Carelink 2090 Firmware, Medtronic Amplia Crt-D Firmware, Medtronic Claria Crt-D Firmware. Version information: version 2490.

RemediationAI

No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Apply vendor patches when available. Implement network segmentation and monitoring as interim mitigations.

Share

CVE-2019-6538 vulnerability details – vuln.today

This site uses cookies essential for authentication and security. No tracking or analytics cookies are used. Privacy Policy