Skip to main content

Sigma Spectrum Infusion System Firmware

6 CVEs product

Monthly

CVE-2020-12047 CRITICAL Act Now

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.

Authentication Bypass Sigma Spectrum Infusion System Firmware
NVD
CVSS 3.1
9.8
EPSS
1.7%
CVE-2020-12045 CRITICAL Act Now

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.

Authentication Bypass Sigma Spectrum Infusion System Firmware
NVD
CVSS 3.1
9.8
EPSS
1.7%
CVE-2020-12043 CRITICAL Act Now

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP service operating on the WBM remains operational until the WBM is rebooted. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.

Information Disclosure Sigma Spectrum Infusion System Firmware
NVD
CVSS 3.1
9.8
EPSS
2.1%
CVE-2020-12041 CRITICAL Act Now

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to. Rated critical severity (CVSS 9.4), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.

Information Disclosure Sigma Spectrum Infusion System Firmware
NVD
CVSS 3.1
9.4
EPSS
1.4%
CVE-2020-12040 CRITICAL Act Now

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.

Information Disclosure Sigma Spectrum Infusion System Firmware
NVD
CVSS 3.1
9.8
EPSS
0.9%
CVE-2020-12039 LOW Monitor

Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered. Rated low severity (CVSS 2.4), this vulnerability is no authentication required, low attack complexity. No vendor patch available.

Authentication Bypass Sigma Spectrum Infusion System Firmware
NVD
CVSS 3.1
2.4
EPSS
0.3%
EPSS 2% CVSS 9.8
CRITICAL Act Now

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.

Authentication Bypass Sigma Spectrum Infusion System Firmware
NVD
EPSS 2% CVSS 9.8
CRITICAL Act Now

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.

Authentication Bypass Sigma Spectrum Infusion System Firmware
NVD
EPSS 2% CVSS 9.8
CRITICAL Act Now

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP service operating on the WBM remains operational until the WBM is rebooted. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.

Information Disclosure Sigma Spectrum Infusion System Firmware
NVD
EPSS 1% CVSS 9.4
CRITICAL Act Now

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to. Rated critical severity (CVSS 9.4), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.

Information Disclosure Sigma Spectrum Infusion System Firmware
NVD
EPSS 1% CVSS 9.8
CRITICAL Act Now

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication. Rated critical severity (CVSS 9.8), this vulnerability is remotely exploitable, no authentication required, low attack complexity. No vendor patch available.

Information Disclosure Sigma Spectrum Infusion System Firmware
NVD
EPSS 0% CVSS 2.4
LOW Monitor

Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered. Rated low severity (CVSS 2.4), this vulnerability is no authentication required, low attack complexity. No vendor patch available.

Authentication Bypass Sigma Spectrum Infusion System Firmware
NVD