Skip to main content

Sc 6002Xl CVE-2019-25722

| EUVDEUVD-2019-20158 HIGH
Use of Hard-coded Credentials (CWE-798)
2026-06-02 VulnCheck GHSA-fh3f-j952-25x6
7.2
CVSS 4.0 · Vendor: VulnCheck
View

Vulnerability Assessment

Weakness
CWE
OWASP Top 10:2025
A07:2025 - Authentication Failures
ASVS 5.0
V6 - Authentication Credential externalization
MITRE ATT&CK

2 techniques across 2 tactics: Persistence, Credential Access

The technique list with ATT&CK links and the kill-chain view is part of the Basic plan and up.

Impact

CVSS score of 7.2 (HIGH). Attack characteristics: adjacent network attack vector, low complexity, no privileges required, no user interaction not required. EPSS exploitation probability: 0.04%.

Exploit ScenarioDerived

An attacker could exploit this vulnerability to gain access using credentials embedded in source code. No authentication is required to exploit this vulnerability.

This Week Schedule remediation this week. Why: HIGH severity (CVSS 7.0-8.9).
No patch listed yet 0.0% exploitation probability (EPSS)

Get an e-mail when this CVE is exploited, patched or re-scored. Free account.

Alert me

AnalysisAI

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and. Rated high severity (CVSS 7.2), this vulnerability is no authentication required, low attack complexity. No vendor patch available.

RemediationAI

No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Remove hard-coded credentials, use environment variables or secrets management, rotate exposed credentials immediately.

Affected VersionsENISA

SC 7000 all versions
SC6802XL all versions
SC 6002XL all versions ≤VG2.3.1
SC90000 XL all versions
SC8000 all versions

Affects your stack

Find out whether the stacks and servers you run are on an affected version.

Checking this CVE against your own stacks and servers is part of the Basic plan and up.

Technical ContextAI · show details

This vulnerability is classified as Use of Hard-coded Credentials (CWE-798), which allows attackers to gain access using credentials embedded in source code. Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring. Affected products include: Dräger Sc 6002Xl, Dräger Sc6802Xl, Dräger Sc 7000, Dräger Sc8000, Dräger Sc90000 Xl.

Know when Sc 6002Xl gets its next vulnerability

  • Watch products and get an email when something new is published
  • Check this CVE against your own stack and servers
  • Exploit and PoC references, threat intelligence from 22 sources

No credit card · 7-day full trial

Severity by source

Vendor (VulnCheck) PRIMARY 7.2 HIGH
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N

Primary rating from Vendor (VulnCheck) · only source for this CVE.

CVSS VectorVendor: VulnCheck

Attack Vector
Adjacent
Attack Complexity
Low
Attack Requirements
None
Privileges Required
None
User Interaction
None
Vulnerable system: Confidentiality
Low
Vulnerable system: Integrity
Low
Vulnerable system: Availability
High
Subsequent system: Confidentiality
None
Subsequent system: Integrity
None
Subsequent system: Availability
None

Lifecycle Timeline

2
CVSS changed
Jun 02, 2026 - 20:22 NVD
7.6 (HIGH) 7.2 (HIGH)
CVE Published
Jun 02, 2026 - 18:44 cve.org
HIGH 7.2

DescriptionCVE.org

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.

Threat Intelligence Overview

EPSS Score
0.0%
Top 87% exploitable
Data Quality
5 sources
Confidence: Medium

How this decision was reached

The engine reads each signal below and applies a fixed list of rules from top to bottom. The first rule that matches sets the recommended action – like a statement, every line is shown.

SignalValue
Severity HIGH
CISA KEV No
Public POC No
EPSS 0.0%
Patch available No
Active threat No
Deciding rule
HIGH severity (CVSS 7.0-8.9)
What every action level means
Emergency Immediate action required. Criteria: CRITICAL + in CISA KEV (actively exploited) without patch, OR CRITICAL + EPSS > 30% without patch, OR CRITICAL + active threat signals without patch.
Act Now Prioritize within 24-48h. Criteria: in KEV with patch available, OR CRITICAL + public POC, OR HIGH + EPSS > 10%, OR CRITICAL + active threat, OR any CRITICAL severity.
This Week Schedule remediation this week. Criteria: HIGH severity (CVSS 7.0-8.9). Patch within normal change window.
This Month Plan within 30 days. Criteria: MEDIUM severity (CVSS 4.0-6.9), lower exploitation probability. Include in regular patch cycle.
Monitor Track and reassess. Criteria: LOW severity (CVSS 0.1-3.9), minimal exploitation risk. No immediate action needed.

SSVC AssessmentENISA

SSVC Decision· derived
Scheduled
Exploitation
None
System Exposure· derived
Controlled
Automatable
No
Human Impact· derived
Medium
Technical Impact
Partial

Priority Score

36
Low Medium High Critical
KEV: 0
EPSS: 0
CVSS: +36
PoC: 0
Weaponized: 0
Threat intel: 0
Gov advisory: 0

Attack Technique

Authentication bypass attacks exploit flaws in the verification mechanisms that control access to systems and applications.

This site uses cookies essential for authentication and security. No tracking or analytics cookies are used. Privacy Policy