Skip to main content

Accu Chek Inform Ii Firmware CVE-2018-18565

MEDIUM
Unrestricted Upload of File with Dangerous Type (CWE-434)
2018-11-20 cve@mitre.org
6.8
CVSS 3.0 · NVD
Share

Severity by source

NVD PRIMARY
6.8 MEDIUM
AV:A/AC:L/PR:L/UI:N/S:C/C:N/I:H/A:N

Primary rating from NVD · only source for this CVE.

CVSS VectorNVD

CVSS:3.0/AV:A/AC:L/PR:L/UI:N/S:C/C:N/I:H/A:N
Attack Vector
Adjacent
Attack Complexity
Low
Privileges Required
Low
User Interaction
None
Scope
Changed
Confidentiality
None
Integrity
High
Availability
None

Lifecycle Timeline

1
CVE Published
Nov 20, 2018 - 19:29 nvd
MEDIUM 6.8

DescriptionNVD

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial number below KQ0400000 or KS0400000), and cobas h 232 before 04.00.04 (Serial number above KQ0400000 or KS0400000). A vulnerability in the software update mechanism allows authenticated attackers in the adjacent network to overwrite arbitrary files on the system through a crafted update package.

AnalysisAI

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00,. Rated medium severity (CVSS 6.8), this vulnerability is low attack complexity. No vendor patch available.

Technical ContextAI

This vulnerability is classified as Unrestricted File Upload (CWE-434), which allows attackers to upload malicious files that can be executed on the server. An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial number below KQ0400000 or KS0400000), and cobas h 232 before 04.00.04 (Serial number above KQ0400000 or KS0400000). A vulnerability in the software update mechanism allows authenticated attackers in the adjacent network to overwrite arbitrary files on the system through a crafted update package. Affected products include: Roche Accu-Chek Inform Ii Firmware, Roche Cobas H 232 Firmware, Roche Coaguchek Pro Ii Firmware, Roche Coaguchek Xs Plus Firmware, Roche Coaguchek Xs Pro Firmware. Version information: before 03.06.00.

RemediationAI

No vendor patch is available at time of analysis. Monitor vendor advisories for updates. Validate file types server-side, store uploads outside webroot, use random filenames, scan for malware.